Página 1 - CardioSys/MicroLab

CardioSys/MicroLabServicing Instructions227 436 37 SA(e) Revision BVersion 4.x

Página 2

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 10Servicing Instructions 227 436 37 Rev. B V1.33.1.5 Internal power distributerThe syst

Página 3 - Table of Contents

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 100Servicing Instructions 227 436 37 Rev. B V1.32001967-001SCSI Controller Adaptec AVA-29

Página 4

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 101Servicing Instructions 227 436 37 Rev. B V1.3MicroLab Addition (202 316 ..)93011798 P

Página 5

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 102Servicing Instructions 227 436 37 Rev. B V1.391920155 Mains Extension with 6 Outputs91

Página 6

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 103Servicing Instructions 227 436 37 Rev. B V1.375130018 Thumbsrew M550465745 Mounting An

Página 7 - 1 General information

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 104Servicing Instructions 227 436 37 Rev. B V1.3

Página 8 - 2 Block diagram

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 105Servicing Instructions 227 436 37 Rev. B V1.313 AppendixData Sheets and User’s Manuals

Página 9 - 3 Functional description

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 106Servicing Instructions 227 436 37 Rev. B V1.3Multi I/O Interface Card HandbookUser&apo

Página 10 - 3.2 Computer subassembly

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 107Servicing Instructions 227 436 37 Rev. B V1.3Standard setting ex-factory:Figure 1: Jum

Página 11 - 3.2.7 Floppy

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 108Servicing Instructions 227 436 37 Rev. B V1.3NOTES:Although the system architecture of

Página 12 - 3.2.11 MO drive

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 109Servicing Instructions 227 436 37 Rev. B V1.3Compatible with ISA and EISA systemsMaxim

Página 13

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 11Servicing Instructions 227 436 37 Rev. B V1.3After switching the Power on/off switch to

Página 15 - 3.3.2 Function key keyboard

Marquette Hellige GmbH CardioSys / MicroLab V4.X AppendixServicing Instructions 227 436 37 Rev. AV1.3AppendixWiring DiagramsNetwork-Cards:SE440B

Página 23 - 5 Installation

SE440BXMotherboardProduct GuideOrder Number: 697967-001

Página 24

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 12Servicing Instructions 227 436 37 Rev. B V1.33.2.8 Free slots for upgradingThe free s

Página 25

Revision HistoryRevision Revision History Date-001 First release of the SE440BX Motherboard Product Guide. February, 1998If an FCC declaration of conf

Página 26 - 5.2.7 Entries in WIN.INI

iiiContents1 Motherboard FeaturesFeatures Summary...

Página 27 - [EGM_RECORD]

ContentsivHow to Install Memory...27How to Remov

Página 28

SE440BX Motherboard Product GuidevA Error MessagesBIOS Beep Codes...

Página 29

Contentsvi14. Security Menu ...4115. Power Me

Página 30

71 Motherboard FeaturesThis chapter gives an overview of the SE440BX motherboard, including:• Features• Components• Back panel I/O connectorsThe re

Página 31 - 5.2.8 Entries in AR200.INI

Motherboard Features8ComponentsFigure 1 shows the major components on the motherboard.OM07065B HEDF GCAJIKLUTR Q NOS P MXVWA Wake on Ring connector M

Página 32

SE440BX Motherboard Product Guide9Back Panel I/O ConnectorsFigure 2 shows the back panel I/O connectors on the motherboard.OM06985BD EHIKJAGFCA PS/2†

Página 33

Motherboard Features10MicroprocessorThe motherboard supports a single Pentium II processor operating at any of the Pentium IIprocessor speeds, voltage

Página 34

SE440BX Motherboard Product Guide11Input/Output (I/O) ControllerThe I/O controller handles the exchange of information between the processor and exter

Página 35

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 13Servicing Instructions 227 436 37 Rev. B V1.3Formatting Optical DisksYou must have syst

Página 36

Motherboard Features12• Support for up to 127 physical devices• Guaranteed bandwidth and low latencies appropriate for telephony, audio, and otherap

Página 37 - MicroLab Systems

SE440BX Motherboard Product Guide13IDE Auto ConfigurationIf you install an IDE device (e.g., a hard drive) in your computer, the IDE auto-configuratio

Página 38

Motherboard Features14Power ManagementThe motherboard supports two types of power management — Advanced Power Management(APM) and Advanced Configurati

Página 39

SE440BX Motherboard Product Guide15Wake on RingWake on Ring enables the computer to wake from sleep or soft-off mode when a call is receivedon a telep

Página 40 - CardioSys / MicroLab Systems

Motherboard Features16Crystal Semiconductor CS4236B Audio CodecThe CS4236B audio codec’s features include:• Compatibility with Sound Blaster†, Sound B

Página 41

172 Installing and Replacing MotherboardComponentsThis chapter describes the following:• How to install and remove the motherboard• How to install a

Página 42

Installing and Replacing Motherboard Components18How to Install and Remove the MotherboardRefer to your chassis manual for instructions on installing

Página 43

SE440BX Motherboard Product Guide19Install the Retention MechanismTo install the retention mechanism, follow these steps:1. Observe the precautions i

Página 44

Installing and Replacing Motherboard Components20Install the ProcessorTo install the processor, follow these steps:1. Insert the processor in the ret

Página 45

SE440BX Motherboard Product Guide213. Slide the top heatsink support bar onto the retaining pins of the support’s base as shown inFigure 6.AOM07060BA

Página 46

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 14Servicing Instructions 227 436 37 Rev. B V1.33.2.12 Streamer DATAn optional DAT strea

Página 47

Installing and Replacing Motherboard Components22 CAUTIONTo avoid bending or breaking pins, use caution when removing or installing a jumper.To set t

Página 48

SE440BX Motherboard Product Guide23How to Upgrade to a Boxed Pentium® II ProcessorUse the instructions in this section to prepare the motherboard for

Página 49

Installing and Replacing Motherboard Components246. Remove the top bar of the heatsink support from the base as shown in Figure 8. Press in on thela

Página 50

SE440BX Motherboard Product Guide25Remove the Heatsink Support Base✏NOTETo remove the heatsink support base from the motherboard, you need a special r

Página 51

Installing and Replacing Motherboard Components262. Place the heatsink support removal tool over the two outside posts of the heatsink support baseas

Página 52

SE440BX Motherboard Product Guide27Upgrade the ProcessorSee the documentation that came with the boxed Intel Pentium II processor.How to Install Memor

Página 53 - 5.2.15 Option keys

Installing and Replacing Motherboard Components28Figure 12 shows the location of the DIMM sockets.OM07067210Figure 12. Location of DIMM SocketsTo ins

Página 54

SE440BX Motherboard Product Guide29OM07068Figure 13. Installing a DIMMHow to Remove MemoryTo remove a DIMM, follow these steps:1. Observe the precau

Página 55

Installing and Replacing Motherboard Components30How to Replace the BatteryWhen your computer is turned off, a lithium battery maintains the current t

Página 56

SE440BX Motherboard Product Guide31To replace the battery, follow these steps:1. Observe the precautions in “Before You Begin” (see page 17).2. Turn

Página 57 - 5.2.17 PCI PC Bios

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 15Servicing Instructions 227 436 37 Rev. B V1.33.3.2 Function key keyboardThe function

Página 58

Installing and Replacing Motherboard Components32How to Clear the PasswordsThis procedure assumes that the motherboard is installed in the computer an

Página 59 - 6 Instrument options

333 Using the Setup ProgramThis chapter provides an overview of the Setup program. You can use the Setup program tochange the configuration informati

Página 60 - 6.1.9 Connector cable T2000

Using the Setup Program34Setup MenusTo enter the Setup program, turn the computer on and press <F2> when you see the message:Press <F2> Ke

Página 61 - 6.2 Software options

SE440BX Motherboard Product Guide35Maintenance MenuThis menu is for setting the processor speed and clearing the Setup passwords. Setup only displays

Página 62

Using the Setup Program36Advanced MenuThis menu is for setting advanced features that are available through the chipset.Table 6. Advanced MenuFeature

Página 63 - 7 Instrument versions

SE440BX Motherboard Product Guide37Peripheral Configuration SubmenuThis submenu is for the configuring the computer peripherals.Table 7. Peripheral Co

Página 64

Using the Setup Program38Table 7. Peripheral Configuration Submenu (continued)Feature Options DescriptionParallel port • Disabled• Enabled• Auto (d

Página 65 - 8 Troubleshooting tips

SE440BX Motherboard Product Guide39IDE Configuration SubmenusThis submenu is for configuring IDE devices, including:• Primary IDE master• Primary ID

Página 66

Using the Setup Program40Floppy OptionsThis submenu is for configuring floppy drives.Table 10. Floppy OptionsFeature Options DescriptionFloppy Disk Co

Página 67 - 8.3 Printer debugging

SE440BX Motherboard Product Guide41Resource Configuration SubmenuThis submenu is for configuring the memory and interrupts.Table 13. Resource Configur

Página 68

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 16Servicing Instructions 227 436 37 Rev. B V1.33.3.4 MonitorA high-resolution 15" o

Página 69

Using the Setup Program42Power MenuThis menu is for setting power management features.Table 15. Power MenuFeature Options DescriptionPower Management

Página 70

SE440BX Motherboard Product Guide43Table 16. Boot Menu (continued)Feature Options DescriptionFirst Boot DeviceSecond Boot DeviceThird Boot DeviceFourt

Página 71

Using the Setup Program44Removable Devices SubmenuThis submenu is for configuring the boot sequence for removable devices.Table 18. Removable Devices

Página 72

454 Upgrading the BIOSThis chapter describes how to upgrade the BIOS and how to recover the BIOS if an upgrade fails.Preparing for the UpgradeBefore y

Página 73

Upgrading the BIOS46Creating a Bootable Diskette✏NOTEIf your drive A is an LS-120 diskette drive, you must use a 1.44-MB diskette as the bootable BIOS

Página 74

SE440BX Motherboard Product Guide475. When the utility asks for confirmation that you want to flash the new BIOS into memory,select Continue with Pro

Página 75

Upgrading the BIOS4810. On the jumper block (J8A1), move the jumper back to pins 1-2 as shown below to set normalmode for Setup.13J8A1OM0707211. Leave

Página 76

495 Technical ReferenceMotherboard ConnectorsFigure 15 shows the location of some of the motherboard connectors. OM07066111111111111020KL J IAEB FCDG

Página 77

Technical Reference50Table 20. Wake on Ring ConnectorPin Signal Name1 Ground2 RINGA#Table 21. Wake on LAN TechnologyConnectorPin Signal Name1 +5 VSB2

Página 78 - 8.5 General Troubleshooting

SE440BX Motherboard Product Guide51Table 25. CD Audio ConnectorPin Signal Name1 CD_IN-Left2 Ground3 Ground4 CD_IN-RightTable 26. Chassis Intrusion C

Página 79

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 17Servicing Instructions 227 436 37 Rev. B V1.3* PS/2 mouse interface* keyboard interface

Página 80

Technical Reference52Table 29. Power Supply ConnectorPin Signal Name Pin Signal Name1 +3.3 V 11 +3.3 V2 +3.3 V 12 -12 V3 Ground 13 Ground4 +5 V 14 PS-

Página 81

SE440BX Motherboard Product Guide53Front Panel ConnectorsThe motherboard has connectors for controls and indicators typically located on the front pan

Página 82

Technical Reference54Motherboard ResourcesMemory MapTable 33. Memory MapAddress Range (decimal) Address Range (hex) Size Description1024 K - 393216 K

Página 83 - 9 Care and maintenance

SE440BX Motherboard Product Guide55I/O MapTable 35. I/O MapAddress (hex) Size Description0000 - 000F 16 bytes DMA controller 10020 - 0021 2 bytes Inte

Página 84 - 9.1.2 Test functions

Technical Reference56PCI Configuration Space MapTable 36. PCI Configuration Space MapBusNumber (hex)DeviceNumber (hex)FunctionNumber (hex) Description

Página 85 - 9.2 Safety Analysis Test

57A Error MessagesBIOS Beep CodesOne long beep followed by several short beeps indicates a video problem.Table 38. Beep CodesBeeps Port 80h Code Expla

Página 86

Error Messages58Table 39. BIOS Error Messages (continued)Error Message ExplanationFixed Disk 0 Failure orFixed Disk 1 Failure orFixed Disk Controller

Página 87

59B Regulatory and Integration InformationThis appendix contains:• Safety standards, electromagnetic compatibility regulations, and product certifica

Página 88

Regulatory and Integration Information60Product Certification MarkingsThis printed circuit board assembly has the following product certification mark

Página 89 - 10 Jumper tables

SE440BX Motherboard Product Guide61Ensure Electromagnetic Compatibility (EMC)Before computer integration, make sure that the power supply and other mo

Página 90 - 10.4 Network Card

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 18Servicing Instructions 227 436 37 Rev. B V1.33.4.3 Mouse/trackball interfaceThe mouse

Página 91 - 10.5 SCSI Card

Regulatory and Integration Information62Prevent Power Supply OverloadUnless the power supply has inherent overcurrent protection, do not overload the

Página 93

Matrox Graphics ArchitectureMatrox Graphics Card• Installation Guide •10526-101-03001998.02.06ENGLISH

Página 94

2 Matrox Graphics Card – Installation GuideContentsUsing this guide 3Overview 4Hardware installation 5Software installation 7Software setup – Windows

Página 95 - 11 Specifications

Matrox Graphics Card – Installation Guide 3Using this guideThis guide assumes you’re familiar with basic functions like click, right-click and double-

Página 96

4 Matrox Graphics Card – Installation GuideOverviewThank you for purchasing a Matrox graphics card. Your Matrox Millennium II, Mystique 220, or G100 s

Página 97 - 12 Spare parts list

Matrox Graphics Card – Installation Guide 5Hardware installation1Switch to the standard VGA display driverIf you’re running Windows 95 on your compute

Página 98 - MicroLab

6 Matrox Graphics Card – Installation Guide4Insert your Matrox cardRemove the cover for the slot you intend to use. Save the screw for the mounting br

Página 99 - Component's

Matrox Graphics Card – Installation Guide 7Software installationThis section describes how to install Matrox software for Windows 95 and WindowsNT4.0.

Página 100 - Software

8 Matrox Graphics Card – Installation GuideSoftware setup – Windows 95To access PowerDesk programs and online documentation, click  Programs  MGA

Página 101

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 19Servicing Instructions 227 436 37 Rev. B V1.33.4.7 CORINA interfaceThe CORINA interfa

Página 102

Matrox Graphics Card – Installation Guide 9Software overviewMGA PowerDesk for Windows 95 includes Display driver with: DirectDraw, DirectVideo and D

Página 103

10 Matrox Graphics Card – Installation GuideSoftware setup – Windows NT 4.0To access PowerDesk programs and online documentation, click  Programs 

Página 104

Matrox Graphics Card – Installation Guide 11Software overviewMGA PowerDesk for Windows NT 4.0 includes Display driver with: Multi-display support to

Página 105 - 13 Appendix

12 Matrox Graphics Card – Installation GuideTroubleshootingThis section contains solutions to problems you may encounter with your Matrox graphics car

Página 106

Matrox Graphics Card – Installation Guide 13Problem: After the startup screen, or after display settings are changed,the screen image is garbled or un

Página 107

14 Matrox Graphics Card – Installation GuideProblem: After game for Windows 95 starts, monitor doesn’t display properly(rolling screen images, overlap

Página 108

Matrox Graphics Card – Installation Guide 15Cause(PCI graphics cards only) Your Matrox card may be in a “slave” expansion slot on a computer that make

Página 109

16 Matrox Graphics Card – Installation GuideProblem: Game for Windows 95 doesn’t start or runs slower than normal(program uses Microsoft DirectX inter

Página 110

Matrox Graphics Card – Installation Guide 17SolutionIf possible, update your Matrox display driver. Matrox makes new display drivers available on the

Página 111 - Appendix

18 Matrox Graphics Card – Installation GuideProblem: Screen image defects appear in Windows 95 or Windows NT 4.0(example: mouse pointer not drawn prop

Página 112

Caution:During repairs/service interventions,observe the protective measures againstdamage due to ESD.* Marquette Hellige GmbH considers itselfrespons

Página 113

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 20Servicing Instructions 227 436 37 Rev. B V1.33.4.8 Printer portThe printer port is on

Página 114

Matrox Graphics Card – Installation Guide 19Display informationRefresh ratesMaximum display resolutions – Matrox Mystique 220 & G100 seriesDisplay

Página 115

20 Matrox Graphics Card – Installation GuideMaximum display resolutions – Matrox Millennium IISupported VESA modesIn the table below, VESA modes suppo

Página 116

Matrox Graphics Card – Installation Guide 21Hardware informationMatrox G100 series MGA-G100 graphics chip, 230 MHz internal RAMDAC Graphics cards –

Página 117

22 Matrox Graphics Card – Installation GuideMatrox G100 series, Millennium II & Mystique 220 Technical features – 64-bit VGA-compatible drawing

Página 118

Matrox Graphics Card – Installation Guide 23Customer supportMatrox Web and FTP sitesMatrox is on the Internet with a World Wide Web (WWW) and File Tra

Página 119 - Product Guide

24 Matrox Graphics Card – Installation Guide5 Direct Matrox Technical Assistance – If you still can’t resolve a problem, you can get direct technical

Página 120 - Revision History

Matrox Graphics Card – Installation Guide 25Free CompuServe introductory membership offerAs a Matrox customer, you are entitled to a free Introductory

Página 121 - Contents

26WarrantyMatrox hardware products are warranted against defects in materials and workmanship for a period of three years from the date of delivery. W

Página 122 - 5 Technical Reference

Matrox Graphics Card – Installation Guide 27Index0-93D4, 14Aacceleration, MGA16, 18address, Matrox24Adobe Acrobat3AGP5, 22BBBS, Matrox23BIOS settings1

Página 123 - A Error Messages

28 Matrox Graphics Card – Installation GuideSoftware license agreementSingle User Products This Matrox software (the “Software”) is copyrighted by Mat

Página 124

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 21Servicing Instructions 227 436 37 Rev. B V1.34 System/Connector Configuration4.1 Pen

Página 125 - 1 Motherboard Features

Copyright © 1998 Matrox Graphics Inc. • (English) All rights reserved. • (Français) Tous droits réservés. • (Deutsch) Alle Rechte vorbehalten. • (Ital

Página 126 - Components

(English) DisclaimerTHE INFORMATION IN THIS GUIDE IS SUBJECT TO CHANGE AT ANY TIME AND WITHOUT NOTICE.Matrox Graphics Inc. reserves the right to make

Página 128 - PCI Enhanced IDE Interface

Excerpt ofonly hardware and softwareinstallation

Página 135

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 22Servicing Instructions 227 436 37 Rev. B V1.3PC top viewConnector view

Página 146

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 23Servicing Instructions 227 436 37 Rev. B V1.35 Installation5.1 Installation of the h

Página 153 - Main Menu

RInstallationGuideAHA-2910CPCI-to-Fast SCSI Host AdapterAdaptec, Inc.691 South Milpitas Blvd.Milpitas, CA 95035© 1997, Adaptec, Inc. All rights reserv

Página 154 - Advanced Menu

1IntroductionThis installation guide provides step-by-step instructions on install-ing your AHA®-2910C host adapter. This process involves installing

Página 155 - ATI mach64

2InstallationInstalling the Host AdapterStep 1: Turn OFF your computer and attached devices, and then disconnect the power cords.Step 2: Touch the met

Página 156 - IDE Configuration

3Step 6: OPTIONAL. If you want the LED in your computer to light whenever there is activity on the SCSI bus, disconnect the LED cable from the LED con

Página 157 - IDE Configuration Submenus

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 24Servicing Instructions 227 436 37 Rev. B V1.3* Patient cableConnect up patient cable to

Página 158 - Video Configuration Submenu

4Step 3: Align the SCSI cable’s colored stripe with Pin-1 on the SCSI device connector, and then plug it in. Step 4: Attach the remaining devices usin

Página 159 - Security Menu

5Step 4: Reconnect the power cables to the computer and external SCSI device. Step 5: Turn ON the external device and computer. Step 6: Go to Installi

Página 160 - Boot Menu

6 NetWare, OS/2, and UNIX: Drivers for these operat-ing systems are not embedded, so you must load them yourself. If your host adapter came bundled wi

Página 161 - Hard Drive Submenu

7If you connected a new SCSI hard disk drive to your host adapter, you must partition and logically format the drive. For DOS and Windows (3.x and 95)

Página 162 - Exit Menu

8Default SettingsThe AHA-2910C has default settings appropriate for most PCI systems (see the table of settings below). Do not run SCSISelect unless y

Página 163

9Using the SCSI Disk UtilitiesTo access the SCSI disk utilities, select the SCSI Disk Utilities option from the menu that appears after start-ing SCSI

Página 164 - Upgrading the BIOS

10Common Problems and Solutions Changed Values Not Loaded—If you changed any values on the host adapter in a Setup program or on a SCSI device, select

Página 165 - Recovering the BIOS

11Federal Communications Commission Radio Frequency Interference StatementWARNING: Changes or modifications to this unit not expressly approved by the

Página 167

CORINAServicing InstructionsVersion 2.0101 118 01...24227 437 01 SA(e) Revision D

Página 168

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 25Servicing Instructions 227 436 37 Rev. B V1.35.2.1 Installation of CardioSoftFor Card

Página 169 - Table 28. Fan 1 Connector

Caution:During repairs/service interventions,observe the protective measures againstdamage due to ESD.* Marquette Hellige GmbH considers itselfrespons

Página 170 - Table 31. PC/PCI Connector

Marquette Hellige GmbH CORINA Page 3Version 2.0 227 437 01 DContents1. Context... ...

Página 171 - Front Panel Connectors

Marquette Hellige GmbH CORINA Page 4Version 2.0 227 437 01 DRevision HistoryEach page of this manual has the document number followed by a revision le

Página 172 - DMA Channels

Marquette Hellige GmbH CORINA Page 5Version 2.0 227 437 01 D1. ContextCORINA is an ECG recorder for the system CardioSys and CardioSoft. CORINA stands

Página 173 - Table 35. I/O Map

Marquette Hellige GmbH CORINA Page 6Version 2.0 227 437 01 D1.4 Model 4 CORINA with Suction Pump and Analog Outputis a combination of Model 2 plus add

Página 174 - Interrupts

Marquette Hellige GmbH CORINA Page 7Version 2.0 227 437 01 D1.7 Replacement of CORINA Model 1...4 through newerCORINA Variants with enhanced communica

Página 175

Marquette Hellige GmbH CORINA Page 8Version 2.0 227 437 01 D

Página 176 - (continued)

Marquette Hellige GmbH CORINA Page 9Version 2.0 227 437 01 D2. IntroductionThe “Hardware Design Description“ describes the structure of the hardware i

Página 177 - Regulatory Compliance

Marquette Hellige GmbH CORINA Page 10Version 2.0 227 437 01 D

Página 178 - Installation Instructions

Marquette Hellige GmbH CORINA Page 11Version 2.0 227 437 01 D3. Hardware Structure3.1 Physicomechanical StructureCORINA comprises 4 (optionally 8) mec

Página 179

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 26Servicing Instructions 227 436 37 Rev. B V1.3Installation Steps:1. Install CardioSoft V

Página 180 - Prevent Power Supply Overload

Marquette Hellige GmbH CORINA Page 12Version 2.0 227 437 01 DUpper case shell with integrated PCB, incl. PCB CORINA TRIGGER:PumpAnalog3.1.2 Floating s

Página 181

Marquette Hellige GmbH CORINA Page 13Version 2.0 227 437 01 DLower case shell with integrated pump module (new tubing system):

Página 182 - • Installation Guide •

Marquette Hellige GmbH CORINA Page 14Version 2.0 227 437 01 D3.1.3 PCB CORINAThere are 5 PCB versions:a) 2x Standard b) 2x for Analog Out c) Standalon

Página 183

Marquette Hellige GmbH CORINA Page 15Version 2.0 227 437 01 D3.1.4 Cable to PC:The connection cable to the PC has a 25-pin plug on the PC side“. The p

Página 184 - Using this guide

Marquette Hellige GmbH CORINA Page 16Version 2.0 227 437 01 DPin Connections (3-Pin Plug)Pin Wire Color Pin Wire Color Pin Wire Color1rd 2 bror ye(*)

Página 185 - Overview

Marquette Hellige GmbH CORINA Page 17Version 2.0 227 437 01 DPump:When using the PCB Analog Out the power line to the suction pump (optional) is notpl

Página 186 - Hardware installation

Marquette Hellige GmbH CORINA Page 18Version 2.0 227 437 01 D3.1.7 Insulation Foil for PCB CORINA TRIGGERThe insulation foil serves to effect floating

Página 187 - Install the software

Marquette Hellige GmbH CORINA Page 19Version 2.0 227 437 01 D3.2.2 CONTROLLER:- all CORINA control functions- initialization of hardware- data communi

Página 188 - Software installation

Marquette Hellige GmbH CORINA Page 20Version 2.0 227 437 01 DMemory:A 128 kByte FLASH ( 128 K x 8 PEROM ), which can be programmed directly with the5-

Página 189 - Software setup – Windows 95

Marquette Hellige GmbH CORINA Page 21Version 2.0 227 437 01 Dvoltage from the 12 V received from the plug-in power supply. The DC/DC transducer canbe

Página 190 - Software overview

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 27Servicing Instructions 227 436 37 Rev. B V1.3GRA_PrinterFont1 = “...” // Default =”Time

Página 191 - Display setup

Marquette Hellige GmbH CORINA Page 22Version 2.0 227 437 01 DSuitable low-pass configuration:2K1uF3.3 Internal Interfaces3.3.1 Mechanical InterfacesTu

Página 192

Marquette Hellige GmbH CORINA Page 23Version 2.0 227 437 01 DInterface to pump module:Description of interface:The pump module interface is in the for

Página 193 - Troubleshooting

Marquette Hellige GmbH CORINA Page 24Version 2.0 227 437 01 D3.4.2 Electrical InterfacesCORINA has two external interfaces. These are:- interface to P

Página 194 -  Settings tab

Marquette Hellige GmbH CORINA Page 25Version 2.0 227 437 01 DPower source interface:As of 05/1997 this interface no longer delivers a 5-V signal.3.4.2

Página 195

Marquette Hellige GmbH CORINA Page 26Version 2.0 227 437 01 D3.4.2.3 Analog Out213ECGGroundPin assignment of this socket: Pin 1 = Analog Out, Pin 2 =

Página 196 -  Programs  MGA PowerDesk 

Marquette Hellige GmbH CORINA Page 27Version 2.0 227 437 01 D4. Troubleshooting Help FunctionsAfter startup, CORINA performs an internal self-test. Th

Página 197

Marquette Hellige GmbH CORINA Page 28Version 2.0 227 437 01 D

Página 198

Marquette Hellige GmbH CORINA Page 29Version 2.0 227 437 01 D5. Safety Analysis Test5.1 General introductionThe suggested Safety Analysis Tests refer

Página 199

Marquette Hellige GmbH CORINA Page 30Version 2.0 227 437 01 D5.1.2.1 Enclosure Leakage Current TestThis test is performed to measure leakage current f

Página 200 - Display information

Marquette Hellige GmbH CORINA Page 31Version 2.0 227 437 01 D5.1.2.2 Patient Leakage Current TestThis test performs a leakage current test under singl

Página 201 - Supported VESA modes

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 28Servicing Instructions 227 436 37 Rev. B V1.3CardioSys updates every 5 seconds the shar

Página 202 - Hardware information

Marquette Hellige GmbH CORINA Page 32Version 2.0 227 437 01 D5.1.2.3 Enclosure Leakage Current Test (System)IMPORTANTDo not operate devices (PC / VGA

Página 203 - Monitor connector pinouts

Marquette Hellige GmbH CORINA Page 33Version 2.0 227 437 01 D6. Technical SpecificationsAnalog output dataImportant: ECG signal output to synchronize

Página 204 - Customer support

Marquette Hellige GmbH CORINA Page 34Version 2.0 227 437 01 DDimensions and Weightlength 400 mmdepth 200 mmheight 48 mmweight w/o. pump700 gwith pump

Página 205 - Where to get information

Marquette Hellige GmbH CORINA Page 35Version 2.0 227 437 01 D7. Spare Part ListShort Description:CORINA is an ECG recorder for the system CardioSys /

Página 206

Marquette Hellige GmbH CORINA Page 36Version 2.0 227 437 01 D38401747 Standard Service Kit Corina Analogout (for programmingof exch. pcb necessary) (c

Página 207 - Warranty

Marquette Hellige GmbH CORINA Page 37Version 2.0 227 437 01 DMounting Details 101 118 21...248. Reference Drawings

Página 208

Marquette Hellige GmbH CORINA Page 38Version 2.0 227 437 01 D

Página 209 - Software license agreement

Marquette Hellige GmbH CORINA Page 39Version 2.0 227 437 01 D9. Appendix: Drawings

Página 210

Marquette Hellige GmbH CORINA Page 40Version 2.0 227 437 01 D

Página 212

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 29Servicing Instructions 227 436 37 Rev. B V1.310. The stage names are fixed as the

Página 216

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx11

Página 217

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx13

Página 218

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx1B

Página 219

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx14

Página 220

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx10

Página 221

Sheet:ofx2x3x4x5Revision-NoNot equipped in Version :Marquette Hellige GmbHD-79007 FreiburgDate / NameISSUEDAPPROVEDDRAWNDate / NameREVISIONSIndexx16

Página 223

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 3Servicing Instructions 227 436 37 Rev. B V1.3Table of Contents1 General information...

Página 224

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 30Servicing Instructions 227 436 37 Rev. B V1.3GRA_CharSet=? // Range of values: 0..4

Página 230

gepr.gez.AENDERUNGENAend. -Nr. Index Datum/NameHELLIGE GMBHPostfach 728D-79007 FreiburgDatum/NameEntw.A03388 032 02 PLPL. CORINA TRIGGERPCB CORINA T

Página 232

Europe RegionMarquette Hellige GmbHA GE Medical Systems CompanyPostfach 60 02 65D-79032 Freiburg • GermanyTel. +49 761 45 43 - 0Fax +49 761 45 43 - 23

Página 233

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 31Servicing Instructions 227 436 37 Rev. B V1.3// Some printer drivers have problems mana

Página 234

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 32Servicing Instructions 227 436 37 Rev. B V1.35.2.9 Entries in SETUP_CM.INIThis file i

Página 235

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 33Servicing Instructions 227 436 37 Rev. B V1.3Serial port B [Enabled]Mode [Normal]Base I

Página 236 - Installation

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 34Servicing Instructions 227 436 37 Rev. B V1.3On LAN: [Stay Off]On PME: [Stay Off]First

Página 237 - Introduction

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 35Servicing Instructions 227 436 37 Rev. B V1.35.2.10.2 Install a Laser PrinterList of

Página 238

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 36Servicing Instructions 227 436 37 Rev. B V1.35.2.10.3 Installation Instruction HP6(M)

Página 239 - Step 2:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 37Servicing Instructions 227 436 37 Rev. B V1.3Installing the Potential Equalization Pin

Página 240 - Connecting One Device

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 38Servicing Instructions 227 436 37 Rev. B V1.32. Switch off the printer and CardioSys /M

Página 241 - Installing AHA-2910C Software

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 39Servicing Instructions 227 436 37 Rev. B V1.35.2.10.4 Installation Instruction EPL570

Página 242 - Helpful Hints

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 4Servicing Instructions 227 436 37 Rev. B V1.35.2 Installation of the software...

Página 243 - Installing Multiple Adapters

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 40Servicing Instructions 227 436 37 Rev. B V1.3Installing the Additional Protective Earth

Página 244 - Default Settings

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 41Servicing Instructions 227 436 37 Rev. B V1.3CardioSys/MicroLab and the printer ground

Página 245 - Troubleshooting Checklist

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 42Servicing Instructions 227 436 37 Rev. B V1.35.2.10.5 Installing the AR200 Printer Sof

Página 246 - Common Problems and Solutions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 43Servicing Instructions 227 436 37 Rev. B V1.3Printer" wizard. Be sure to select a

Página 247 - Canadian Compliance Statement

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 44Servicing Instructions 227 436 37 Rev. B V1.3AR200P THERMAL ARRAY RECORDERWINDOWS 98 DE

Página 248

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 45Servicing Instructions 227 436 37 Rev. B V1.3parallel port you are using for the record

Página 249 - Servicing Instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 46Servicing Instructions 227 436 37 Rev. B V1.3To set the density by using the printer dr

Página 250

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 47Servicing Instructions 227 436 37 Rev. B V1.3contained inside of the AR200P, and will n

Página 251

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 48Servicing Instructions 227 436 37 Rev. B V1.35.2.11 Directory structure and notes on C

Página 252

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 49Servicing Instructions 227 436 37 Rev. B V1.3CardioSys from version 3.0:(some of the di

Página 253 - 1. Context

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 5Servicing Instructions 227 436 37 Rev. B V1.38.4.3 Error messages [3xxx]: Main program..

Página 254 - 1.5 Model 5 CORINA Standalone

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 50Servicing Instructions 227 436 37 Rev. B V1.3Hel_grid.ttf Special FontFiles in Window

Página 255 - CORINA Block Circuit Diagram

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 51Servicing Instructions 227 436 37 Rev. B V1.3INI filesAn initialization file CARDIO.INI

Página 256 - Version 2.0 227 437 01 D

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 52Servicing Instructions 227 436 37 Rev. B V1.38. Tab "WINS"Enter correct IP-ad

Página 257 - 2. Introduction

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 53Servicing Instructions 227 436 37 Rev. B V1.35.2.14 Service screenThe service screen is

Página 258

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 54Servicing Instructions 227 436 37 Rev. B V1.35.2.16 Data transfer to/from MUSEStore Ex

Página 259 - 3. Hardware Structure

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 55Servicing Instructions 227 436 37 Rev. B V1.3Modem/RAS InstallationPrerequisites:RAS ac

Página 260 - 3.1.2 Floating screening:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 56Servicing Instructions 227 436 37 Rev. B V1.39. Connecting Internet Browser to MUSE via

Página 261

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 57Servicing Instructions 227 436 37 Rev. B V1.36. Confirm by "OK"Hint: Some cha

Página 262 - 3.1.3 PCB CORINA

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 58Servicing Instructions 227 436 37 Rev. B V1.3Example:in the file autoexec.batSET HL_SEA

Página 263 - 3.1.4 Cable to PC:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 59Servicing Instructions 227 436 37 Rev. B V1.36 Instrument options6.1 Hardware options6.

Página 264 - 3.1.5 Pump Module:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 6Servicing Instructions 227 436 37 Rev. B V1.3REFERENCESReference Document Title [SA(e)]

Página 265 - 3.1.6 PCB CORINA TRIGGER

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 60Servicing Instructions 227 436 37 Rev. B V1.36.1.6 Function key keyboards220 092 06 Fu

Página 266 - 3.2.1 ECG Processing

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 61Servicing Instructions 227 436 37 Rev. B V1.36.1.10 Connector cable Tram-Rac (MicroLa

Página 267 - 3.2.2 CONTROLLER:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 62Servicing Instructions 227 436 37 Rev. B V1.3

Página 268 - 3.2.3 PC PORT:

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 63Servicing Instructions 227 436 37 Rev. B V1.37 Instrument versions7.1 System cart C

Página 269

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 64Servicing Instructions 227 436 37 Rev. B V1.3

Página 270 - 3.3 Internal Interfaces

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 65Servicing Instructions 227 436 37 Rev. B V1.38 Troubleshooting tips8.1 System Repair

Página 271 - 3.4 Interfaces to Peripherals

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 66Servicing Instructions 227 436 37 Rev. B V1.3For Windows 95/98:With the Registry Editor

Página 272 - 3.4.2 Electrical Interfaces

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 67Servicing Instructions 227 436 37 Rev. B V1.3------------------------------------------

Página 273

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 68Servicing Instructions 227 436 37 Rev. B V1.3Replace the original printer driver with t

Página 274 - 3.4.2.3 Analog Out

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 69Servicing Instructions 227 436 37 Rev. B V1.3[1115] General data base error (No.%s)An e

Página 275

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 7Servicing Instructions 227 436 37 Rev. B V1.31 General informationCardioSys is a complet

Página 276

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 70Servicing Instructions 227 436 37 Rev. B V1.3[1156] Error when decompressing dataThe se

Página 277 - 5. Safety Analysis Test

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 71Servicing Instructions 227 436 37 Rev. B V1.3[2158] No examinations found on data carri

Página 278

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 72Servicing Instructions 227 436 37 Rev. B V1.3[6508] Error when opening signal processin

Página 279

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 73Servicing Instructions 227 436 37 Rev. B V1.3There are permanent and adjustable load pr

Página 280

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 74Servicing Instructions 227 436 37 Rev. B V1.3- Check option code number under General

Página 281 - 6. Technical Specifications

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 75Servicing Instructions 227 436 37 Rev. B V1.3Remedy:- Restore examination data backup.

Página 282

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 76Servicing Instructions 227 436 37 Rev. B V1.3[13236] No examination data are stored on

Página 283 - 7. Spare Part List

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 77Servicing Instructions 227 436 37 Rev. B V1.3[15012] Error during CORINA downloadingsee

Página 284

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 78Servicing Instructions 227 436 37 Rev. B V1.3[17627] Zero adjustment not possiblePRESS

Página 285 - 8. Reference Drawings

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 79Servicing Instructions 227 436 37 Rev. B V1.3CardioSoft normally checks all LPT ports t

Página 286

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 8Servicing Instructions 227 436 37 Rev. B V1.32 Block diagram

Página 287 - 9. Appendix: Drawings

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 80Servicing Instructions 227 436 37 Rev. B V1.3- Blood pressure display cut offIn the ope

Página 288

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 81Servicing Instructions 227 436 37 Rev. B V1.3- Thermal printer AR200 prints 10mm-platea

Página 289

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 82Servicing Instructions 227 436 37 Rev. B V1.3

Página 290

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 83Servicing Instructions 227 436 37 Rev. B V1.39 Care and maintenance9.1 Technical ins

Página 291

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 84Servicing Instructions 227 436 37 Rev. B V1.39.1.2 Test functionsRecommended testing

Página 292

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 85Servicing Instructions 227 436 37 Rev. B V1.3Identification of disconnected electrodesR

Página 293 - Marquette Hellige GmbH

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 86Servicing Instructions 227 436 37 Rev. B V1.39.2.4 Leakage Current MeasurementTo perf

Página 294

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 87Servicing Instructions 227 436 37 Rev. B V1.3Electrical Diagram for Enclosure Leakage C

Página 295

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 88Servicing Instructions 227 436 37 Rev. B V1.3Electrical Diagram for Patient Leakage Cur

Página 296

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 89Servicing Instructions 227 436 37 Rev. B V1.310 Jumper tables10.1 MOTHERBOARD INTEL

Página 297

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 9Servicing Instructions 227 436 37 Rev. B V1.33 Functional description3.1 Electrical c

Página 298

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 90Servicing Instructions 227 436 37 Rev. B V1.310.3 VGA cardAGP bus, PNP Card, no manua

Página 299

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 91Servicing Instructions 227 436 37 Rev. B V1.310.5 SCSI CardPCI bus, PNP Card, no manual

Página 300

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 92Servicing Instructions 227 436 37 Rev. B V1.310.6 TRAMNET Card (optional)AT Bus, no P

Página 301

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 93Servicing Instructions 227 436 37 Rev. B V1.3Microswitch:12345678┌─────────────────┐On

Página 302

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 94Servicing Instructions 227 436 37 Rev. B V1.3

Página 303

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 95Servicing Instructions 227 436 37 Rev. B V1.311 SpecificationsPower supplyFrom the mai

Página 304

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 96Servicing Instructions 227 436 37 Rev. B V1.3Ambient conditionsOperation:* ambient temp

Página 305 - PCB CORINA TRIGGER

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 97Servicing Instructions 227 436 37 Rev. B V1.312 Spare parts listCardioSys/MicroLab (Wi

Página 306

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 98Servicing Instructions 227 436 37 Rev. B V1.322009223 Function key keyboard Right-Heart

Página 307

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 99Servicing Instructions 227 436 37 Rev. B V1.392308100 Fastening Strap 92 X 2,4 mm923096

Comentários a estes Manuais

Sem comentários