Adaptec 2910C - AHA Storage Controller Fast SCSI 10 MBps Instruções de Operação Página 32

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 31
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 32
Servicing Instructions 227 436 37 Rev. B V1.3
5.2.9 Entries in SETUP_CM.INI
This file is located in the program directory.
Option to set the sorting sequence of the alphanumerical key as required. Example: sort
patients in reverse order; max. 255 characters permitted
[DATABASE]
SortTable=”ZzYyXxWwVvUuTtSsRrQqPpOoNnMmLlKkJjIiHhGgFfEeDdCcBbAa”
This entry only comes into effect when made before calling up CardioSoft for the first
time. The sorting sequence is stored in the database during its creation. If a change in
the sorting sequence is to be made at some later time, the entire database has to be
deleted.
5.2.10 System control settings / printer
5.2.10.1 Bios Configuration
During startup press F2 on request to enter the setup menu. You can leave each menu
by pressing ESC. Changes will be stored.
[Main]
Language: [English (US)]
L2 cache ECC Support: [Disabled]
System Time [Please fill in actual time]
System Date [Please fill in actual date]
[Advanced]
Plug and Play O/S: [No]
Reset Configuration Data [No]
Numlock: [Auto]
[Advanced - Peripheral Configuration]
Serial port A: [Enabled]
Base IO address: [3F8]
Interrupt: [IRQ 4]
Vista de página 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 306 307

Comentários a estes Manuais

Sem comentários