Adaptec 2910C - AHA Storage Controller Fast SCSI 10 MBps Instruções de Operação Página 41

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 40
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 41
Servicing Instructions 227 436 37 Rev. B V1.3
CardioSys/MicroLab and the printer ground strip. The value must be < 0.2
! Also
measure the enclosure leakage current.
8. Connect the printer to a wall outlet and switch it on.
9. If all measurement results are within the admissible range, replace the sticker
"Important: Do not operate the device in the vicinity of the patient (1.5 m/5 ft.)" with the
sticker "CardioSys/MicroLab ..." (refer to Labelling Sheet).
Figure 2
Vista de página 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 306 307

Comentários a estes Manuais

Sem comentários